Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-51
TABLE POINT 01 USER VAL:
CURRENTLY: 0.00
SET TO: 00000.0
0
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Temperature Compensation screen allows the user to turn temperature
compensation on or off. Temperature compensation changes to the
instrument's output signal that occur due to internal instrument
temperature variations. The effects of temperature changes on the analyzer's
subsystems and output have been empirically determined. This empirical
data is used to compensate for any changes in temperature.
When temperature compensation is on, the display shows the current
internal instrument temperature (measured by a thermistor on the Interface
board). When temperature compensation is off, the display shows the
factory standard temperature of 30.0 °C.
●
In the Main Menu, choose Instrument Controls >
Temperature
Compensation
.
TEMPERATURE COMPENSATION:
COMP TEMP: 25.0
o
C
CURRENTLY: ON
SET TO: OFF ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Pressure compensation provides compensation for any changes to the
instrument's output signal due to reaction chamber pressure variations. The
effects of reaction chamber pressure changes on the analyzer's subsystems
and output have been empirically determined. This empirical data is used
to compensate for any change in reaction chamber pressure.
When pressure compensation is on, the first line of the display represents
the current pressure in the fluorescence chamber. When pressure
compensation is off, the first line of the display shows the factory standard
pressure of 750.0 mmHg.
●
In the Main Menu, choose Instrument Controls >
Pressure
Compensation
.
Temperature
Compensation
Pressure
Compensation