Operation
Range Menu
3-14
Model 15
i
Instruction Manual
Thermo Fisher Scientific
RANGE:
CURRENTLY: 50.0
SET TO: 100.0 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Table 3–5.
Standard Ranges
ppm
mgm
3
5 10
10 20
20 50
50 100
100 200
200 500
500 1000
1000 2000
2000 5000
5000 7500
C1 C1
C2 C2
C3 C3
C1, C2, and C3 are custom ranges. For more information about custom
ranges, see “
Set Custom Ranges
” below.
The Set Custom Ranges menu lists three custom ranges, which are user-
defined. In the ppm mode, any value between 5 ppm and 5000 ppm can be
specified as a range. In the mg/m
3
mode, any value between 10 mg/m
3
and
7500 mg/m
3
can be specified as a range.
●
In the Main Menu, choose Range >
Set Custom Ranges
.
CUSTOM RANGES:
>CUSTOM RANGE 1 55.6
CUSTOM RANGE 2 75.0
CUSTOM RANGE 3 125.0
RANGE AVG DIAGS ALARM
Set Custom Ranges