C-Link Protocol Commands
Communications Configuration
B-40
Model 15
i
Instruction Manual
Thermo Fisher Scientific
set power up mode
value
This command is used to configure the instrument to power up in the
local/unlocked mode (
value
= 0
)
or the remote/locked mode (
value
= 1),
according to
Table B–11
.
If the instrument is set to power up in the local/remote mode, the keypad
can be used to make changes via the front panel. If the instrument is set to
power up in the remote/locked mode, changes can not be made from the
front panel. The following example sets the instrument to power up in
remote/locked mode.
Send:
set
power
up
mode
1
Receive:
set power up mode 1 ok
Table B–11.
Power up Mode Values
Value
Power up Mode
0 Local/Unlocked
Mode
(default)
1 Remote/Locked
Mode
program no
This command reports the analyzer’s model information and program
version number.
Send:
program
no
Receive:
program no iSeries 15i 01.05.02.140
layout ack
This command reports the stale layout/layout change indicator (*) that is
attached to each response if the erec layout has changed since the last time
erec layout was requested, according to
Table B–12
. The following example
reports that the instrument is configured to do nothing.
Send:
layout
ack
Receive: layout
ack
0
set layout ack
value
This command disables the stale layout/layout change indicator (*) that is
attached to each response if the erec layout has changed since the last time
erec layout was requested, according to
Table B–12
.
Send:
set
layout
ack
Receive:
set layout ack ok