Servicing
Pressure Transducer Replacement
Thermo Fisher Scientific
Model 15
i
Instruction Manual
7-21
4.
Loosen the four captive screws holding the pump bracket to the shock
mounts and remove the pump assembly and the shock mounts.
5.
Invert the pump assembly, remove the two pump mounting screws
located on the bottom side of the pump bracket, and remove the
bracket from the pump.
6.
Install the new pump by following the previous steps in reverse.
7.
Perform a leak test as described in the “
Preventive Maintenance
”
chapter.
Figure 7–10.
Replacing the Pump
Use the following procedure to replace the pressure transducer (
Figure 7–
11
).
Equipment Required:
Pressure transducer
Philips screwdriver, #2
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
Pressure Transducer
Replacement
Pump
Pump Bracket
Pump Mounting Screws (2)
(Bottom)
Captive Screws (4)
Shock Mounts (4))
Floor Plate