Operation
Instrument Controls Menu
3-40
Model 15
i
Instruction Manual
Thermo Fisher Scientific
I/O CONFIGURATION:
>OUTPUT RELAY SETTINGS
DIGITAL INPUT SETTINGS
ANALOG OUTPUT CONFIG
ANALOG INPUT CONFIG
RANGE AVG DIAGS ALARM
The Output Relay Settings menu displays a list of the 10 digital output
relays available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note
The digital outputs may take up to one second after the assigned
state occurs to show up on the outputs.
▲
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings
.
OUTPUT RELAY SETTINGS:
>1 NOP CONC ALARM
2 NOP LOCAL/REMOTE
3 NOP UNITS
4 NOP GEN ALARM
5 NOP NONE
ª
RANGE AVG DIAGS ALARM
The Logic State menu item is used to change the selected I/O relay to
either normally open or normally closed. The default state is open, which
indicates that a relay connected between the digital output pin and ground
is normally open and closes to trigger the digital output action.
●
Press
to toggle and set the logic state to normally open or
normally closed.
OUTPUT RELAY SETUP:
>LOGIC STATE OPEN
INSTRUMENT STATE
RANGE AVG DIAGS ALARM
Output Relay Settings
Logic State