Servicing
Front Panel Board Replacement
7-40
Model 15
i
Instruction Manual
Thermo Fisher Scientific
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove the three ribbon cables and the two-wire connector from the
front panel board.
3.
Pop the board off of the two top mounting studs and remove the board
by lifting it up and off the slotted bottom support.
4.
Replace the front panel board by following the previous steps in
reverse.
Figure 7–19.
Replacing the Front Panel Board and the LCD Module
Mounting Studs (2)
LCD Module
Retaining Screw
LCD Module Screws (2)
Front Panel Board
Two-Wire Connector
Ribbon Cables (3)
Bottom Support