Calibration
HI and LO Multi-Point Calibration
4-8
Model 15
i
Instruction Manual
Thermo Fisher Scientific
10.
Repeat Step 1 for a 20% concentration of the upper range.
11.
Select Cal Point 3.
12.
Use
to move the cursor and
to increment or
decrement the digit values until they match the concentration being
introduced to the instrument. Press
.
The instrument will perform a series of calculations and save the new
parameters.
13.
Press
to back up a step in the Service Mode menu.
14.
Select Coefficients and press
.
The instrument will automatically calculate the new coefficients and
save the new parameters.
15.
Repeat Steps 1 through 12 for Hi Multi-Point Cal.