![Thermo 15i Instruction Manual Download Page 238](http://html.mh-extra.com/html/thermo/15i/15i_instruction-manual_1104215238.webp)
C-Link Protocol Commands
Datalogging
B-20
Model 15
i
Instruction Manual
Thermo Fisher Scientific
In the following example, assume there are 740 lrecs currently stored in
memory. When the command “lrec 100 5” is sent, the instrument counts
back 100 records from the last record collected (record 740), and then
returns 5 records: 640, 641, 642, 643, and 644. For details on how to
decode the flag fields within these records, see the “flags” command.
Send:
lrec
5
Receive:
lrec 100 5
15:05 08-15-07 flags 8c060000 hcl 7349E+000 hihcl
5994E+000 intt 33.689 cht 44.484 pres 758.886 smplfl
1.085 speed 100.000 biasv -115.883 intensity 199940.000
15:10 08-15-07 flags 8c060000 hcl 7349E+000 hihcl
5994E+000 intt 33.689 cht 44.484 pres 758.886 smplfl
1.085 speed 100.000 biasv -115.883 intensity 199940.000
15:15 08-15-07 flags 8c060000 hcl 7349E+000 hihcl
5994E+000 intt 33.689 cht 44.484 pres 758.886 smplfl
1.085 speed 100.000 biasv -115.883 intensity 199940.000
15:20 08-15-07 flags 8c060000 hcl 7349E+000 hihcl
5994E+000 intt 33.689 cht 44.484 pres 758.886 smplfl
1.085 speed 100.000 biasv -115.883 intensity 199940.000
15:25 08-15-07 flags 8c060000 hcl 7349E+000 hihcl
5994E+000 intt 33.689 cht 44.484 pres 758.886 smplfl
1.085 speed 100.000 biasv -115.883 intensity 199940.000
lrec format
srec format
erec format
These commands report the output format for lrecs and srecs, and erec data
in various formats such as ASCII without text, ASCII with text, or binary.
The following example shows the output format for lrecs is ASCII with
text, according to
Table B–5
.
Send:
lrec
format
Receive: lrec
format
1
set lrec format
format
set srec format
format
set erec format
format
These commands set the output format for lrecs and srecs, and erec data,
according to
Table B–5
. The following example sets the lrec output format
to ASCII with text.
Send:
set lrec format 1
Receive:
set lrec format 1 ok