Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-47
CHOOSE SIGNAL - CONC
SELECTED OUTPUT: V1
CURRENTLY: HCl
SET TO: NONE ?
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Table 3–7.
Signal Type Group Choices
Concentrations
Other Measurements
Analog Inputs
None None None
HCl (singe range mode)
Sample/Reference
Analog Input 1 (if the I/O expansion
board is installed)
LO HCl (dual/auto range
mode)
LO Sample/Reference
Analog Input 2 (if the I/O expansion
board is installed)
HI HCl (dual/auto range
mode)
HI Sample/Reference
Analog Input 3 (if the I/O expansion
board is installed)
Range Status (auto
range mode)
Internal Temp
Analog Input 4 (if the I/O expansion
board is installed)
Bench Temperature
Analog Input 5 (if the I/O expansion
board is installed)
Bench Pressure
Analog Input 6 (if the I/O expansion
board is installed)
Sample Flow
Analog Input 7 (if the I/O expansion
board is installed)
Intensity
Analog Input 8 (if the I/O expansion
board is installed)
Motor
Speed
Bias
Supply
The Analog Input Configuration menu displays a list of the 8 analog input
channels available for configuration. This screen is only displayed if the I/O
expansion board option is installed. Configuration includes entering the
Descriptor, Units, Decimal Places, choice of 1–10 points in the table, and
corresponding number of points selected.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config
.
Analog Input Configuration