Operation
Diagnostics Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-59
ANALOG INPUT VOLTAGES:
>ANALOG IN 1 6.24 V
ANALOG IN 2 4.28 V
ANALOG IN 3 0.00 V
ANALOG IN 4 0.00 V
ANALOG IN 5 0.00 V
ª
RANGE AVG DIAGS ALARM
The Digital Inputs screen (read only) displays the state of the 16 digital
inputs. Pull-ups are provided on all of the inputs, so if nothing they will
read (1), if an input is brought to ground, it will read (0).
●
In the Main Menu, choose Diagnostics >
Digital Inputs
.
DIGITAL INPUTS:
>INPUT 1 1
INPUT 2 1
INPUT 3 1
INPUT 4 1
INPUT 5 1
ª
RANGE AVG DIAGS ALARM
The Relay States screen displays the state of the 10 digital outputs and
allows toggling of the state to either on (1) or off (0). The relays are
restored to their original states upon exiting this screen.
●
In the Main Menu, choose Diagnostics >
Relay States
.
●
Press
to toggle and set the relay state to open or closed for the
selected digital output.
RELAY STATES:
>OUTPUT 1 0
OUTPUT 2 0
OUTPUT 3 0
OUTPUT 4 1
OUTPUT 5 1
ª
RANGE AVG DIAGS ALARM
The Test Analog Outputs menu is used to set each of the analog output
channels to zero or full-scale. Channel choices include all analog outputs, 6
voltage channels, and 6 current channels (if the I/O expansion board
option is installed).
●
In the Main Menu, choose Diagnostics >
Test Analog Outputs
.
Digital Inputs
Relay States
Test Analog Outputs