C-Link Protocol Commands
Hardware Configuration
Thermo Fisher Scientific
Model 15
i
Instruction Manual
B-33
Level
Contrast Level
18 90%
19 95%
20 100%
date
This command reports the current date. The following example reports the
date as December 1, 2006.
Send:
date
Receive: date
12-01-06
set date
mm-dd-yy
mm
= month
dd
= day
yy
= year
This command sets the date of the analyzer’s internal clock. The following
example sets the date to December 1, 2006.
Send:
set
date
12-01-06
Receive:
set date 12-01-06 ok
set default params
This command sets all the parameters to their default values. This does not
affect the factory-calibrated parameters.
Send:
set default params
Receive:
set default params ok
save
set save params
These commands store all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The following example saves the parameters to FLASH
memory.
Send:
set
save
params
Receive:
set save params ok
time
This command reports the current time (24-hour format). The following
example reports that the internal time is 2:15:30 pm.