Servicing
Optional Zero/Span and Sample Solenoid Valve Replacement
Thermo Fisher Scientific
Model 15
i
Instruction Manual
7-27
Capillary
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove the capillary from the inlet elbow fitting on the pump head.
Figure 7–13.
Cleaning or Replacing the Capillary
3.
Clean with less than 0.015-inch diameter wire or replace.
4.
Install the capillary by following the previous steps in reverse.
Use the following procedure to replace the solenoid.
Equipment Required:
Solenoid
Philips screwdriver, #2
Wrench, 9/16-inch
Optional Zero/Span
and Sample Solenoid
Valve Replacement