C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Model 15
i
Instruction Manual
B-45
Send:
list var aout
Receive:
list var aout
index
variable
0
none
5
s/r
8
intt
9
cht
10
auxt
13
pres
14
smplfl
15
intensity
16
speed
17
ain1
18
ain2
19
ain3
20
ain4
21
ain5
22
ain6
23
ain7
24
ain8
28
biasv
29
hcl
relay stat
This command reports the current relay logic as normally “open” or
normally “closed,” if all the relays are set to same state, that is all open or all
closed. The following example shows that the status of all the relays’ logic is
set to normally “open”.
Send:
relay
stat
Receive:
relay stat open
Note
If individual relays have been assigned different logic, then the
response would be a 4-digit hexadecimal string with the least significant
byte (LSB) being relay no 1.
▲
For example:
Receive:
relay stat 0x0001 (indicates relay no 1 is set to
normally open logic, all others are normally closed)
Receive:
relay stat 0x0005 (indicates relay no 1 and 3 are set
to be normally open logic, all others are normally
closed)
set relay open
set relay open
value
set relay closed
set relay closed
value