Operation
Service Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-77
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Analog Out Cal
.
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
ª
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
The Analog Output Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog output. The operator must connect a volt
meter to the output and adjust the output until it reads 0.0 V on the meter
for a voltage channel, or either 0 or 4 mA for a current channel (depending
on the selected range). See the “set output to:” field on the display.
●
In the Main Menu, choose Service > Analog Out Cal > select Channel
>
Calibrate Zero
.
ANALOG OUTPUT CAL: ZERO
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 100
SAVE VALUE
©ª
INC/DEC
SET OUTPUT TO: 0.0 V
RANGE AVG DIAGS ALARM
The Analog Output Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog output. The operator must connect
a volt meter to the output and adjust the output until the meter reads the
value shown in the “set output to:” field, in either V or mA, depending on
the selected output channel.
Analog Output Calibrate
Zero
Analog Output Calibrate
Full-Scale