Servicing
Analog Output Calibration
7-30
Model 15
i
Instruction Manual
Thermo Fisher Scientific
Table 7–4.
Analog Output Channels and Rear Panel Pin Connections
Voltage
Channel
Pin
Current
Channel
Pin
1 14
1
15
2 33
2
17
3 15
3
19
4 34
4
21
5 17
5
23
6 36
6
25
Ground
16, 18, 19, 35, 37
Current Output Return
13, 16, 18, 20, 22, 24
Table 7–5.
Analog Input Channels and Rear Panel Pin Connections
Input Channel
Pin
1 1
2 2
3 3
4 5
5 6
6 7
7 9
8 10
Ground
4, 8, 11, 14
Use the following procedure to calibrate the analog outputs if a meter
reading in the “
Analog Output Testing
” procedure differed by more than
one percent or after replacing the optional I/O expansion board.
Equipment Required:
Multimeter
1.
Connect a meter to the channel to be adjusted and set to voltage or
current as appropriate.
Figure 7–14
shows the analog output pins and
Analog Output
Calibration