Operation
Alarms Menu
3-64
Model 15
i
Instruction Manual
Thermo Fisher Scientific
●
In the Main Menu, choose Alarms >
Pressure
.
PRESSURE:
ACTUAL 746.9 mmHg
>MIN 250.0 mmHg
MAX 1000.0 mmHg
RANGE AVG DIAGS ALARM
The Minimum Pressure alarm limit screen is used to change the minimum
temperature alarm limit. The minimum and maximum pressure screens
function the same way.
●
In the Main Menu, choose Alarms > Pressure >
Min
or
Max
.
PRESSURE:
ACTUAL MIN: 250.0 mmHg
SET MIN TO: 500.0 mmHg?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Sample Flow screen displays the current sample flow reading and sets
the minimum and maximum alarm limits. Acceptable alarm limits range
from .350 to 1.500 LPM. If the sample flow reading goes beyond either the
minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Sample Flow
.
SAMPLE FLOW:
ACTUAL 0.750 LPM
>MIN 0.350 LPM
MAX 1.500 LPM
RANGE AVG DIAGS ALARM
The Minimum Sample Flow alarm limit screen is used to change the
minimum sample flow alarm limit. The minimum and maximum sample
flow screens function the same way.
Min and Max Pressure
Limits
Sample Flow
Min and Max Sample
Flow Limits