Operation
Alarms Menu
3-68
Model 15
i
Instruction Manual
Thermo Fisher Scientific
The Concentration screen displays the current HCl concentration and sets
the minimum and maximum alarm limits. Acceptable alarm limits range
from 0 to 5000 ppm or 0 to 7500 mg/m
3
. The minimum alarm may be
programmed as a floor trigger (alarm is triggered when the concentration
falls below the minimum value) or a ceiling trigger (alarm is triggered when
the concentration goes above the minimum value). If the HCl
concentration goes beyond either the minimum or maximum limit, an
alarm is activated and the alarm (bell) icon appears in the status bar on the
Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Concentration
.
CONCENTRATION:
ACTUAL 58.3
>MIN 0.000
MAX 5000
MIN TRIGGER CEILING
RANGE AVG DIAGS ALARM
The Minimum Concentration alarm limit screen is used to change the
minimum concentration alarm limits. The minimum and maximum
concentration alarm limit screens function the same way.
●
In the Main Menu, choose Alarms > Concentration >
Min
or
Max
.
CONCENTRATION:
ACTUAL MIN: 0.000
SET MIN TO: 000
2
0.00 ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Minimum Trigger screen allows the user to keep set the minimum
concentration alarm trigger type to either floor or ceiling. The minimum
alarm may be programmed as a floor trigger (alarm is triggered when the
concentration falls below the minimum value) or a ceiling trigger (alarm is
triggered when the concentration goes above the minimum value).
●
In the Main Menu, choose Alarms > Concentration >
Min Trigger
.
Concentration
Min and Max
Concentration Limit
Min Trigger
Concentration