C-Link Protocol Commands
Alarms
B-14
Model 15
i
Instruction Manual
Thermo Fisher Scientific
set alarm conc min
value
set alarm conc max
value
These commands set the HCl concentration alarm minimum and
maximum values to
value
, where
value
is a floating-point representation of
the concentration alarm limits. Values must be in the units that are
currently set for use. The following example sets the HCl concentration
alarm maximum value to 5000 ppb.
Send:
set alarm conc max 5000
Receive:
set alarm conc max 5000 ok
alarm internal temp min
alarm internal temp max
These commands report the current internal temperature alarm minimum
and maximum value settings. The following example reports that the
internal temperature alarm minimum value is 15.0 °C.
Send:
alarm internal temp min
Receive:
alarm internal temp min 15.0 deg C
set alarm internal temp min
value
set alarm internal temp max
value
These commands set the internal temperature alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing internal temperature alarm limits in degrees C. The following
example sets the internal temperature alarm maximum value to 45.0 °C.
Send:
set alarm internal temp max 45
Receive:
set alarm internal temp max 45 ok
alarm motor speed min
alarm motor speed max
These commands report the current motor speed alarm minimum and
maximum value settings. The following example reports that the motor
speed alarm minimum value is 99.9%.
Send:
alarm motor speed min
Receive:
alarm motor speed min 99.9
set alarm motor speed min
value
set alarm motor speed max
value
These commands set the motor speed alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing motor
speed alarm limits in percent. The following example below sets the motor
speed alarm maximum value to 100%.
Send:
set alarm motor speed max 100
Receive:
set alarm motor speed max 100 ok