Functional Description
Mass Spectrometer
24
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Figure 11.
Rear power panel of the electronics module
The mass spectrometer main power circuit breaker switch (labeled Main Power) is located on
the mass spectrometer power panel in the lower corner of the right-side panel of the mass
spectrometer. See
. In the Off position, the circuit breaker removes all power to the
mass spectrometer, including the vacuum pumps. In the On position, power is supplied to the
mass spectrometer. In the standard operational mode, the circuit breaker is kept in the On
position.
The electronics service switch (labeled Electronics) is located next to the main power circuit
breaker on the mass spectrometer power panel (
). In the Service Mode position the
switch removes power to all components of the mass spectrometer other than the vacuum
system. The Operating Mode position supplies power to all non-vacuum system components
of the mass spectrometer.
The vacuum service switch (labeled Vacuum) is located next to the electronics service switch
on the mass spectrometer power panel (
). In the Service Mode position the switch
removes power to all components of the vacuum system, including the forepump,
turbomolecular pump, and turbomolecular pump controller. The switch in the Operating
Mode position supplies power to all vacuum system components of the mass spectrometer.
The System Reset button is also located on the mass spectrometer power panel. Pressing the
System Reset button takes the 5 V logic to ground and causes the embedded computer on the
System Control PCB to reboot. The mass spectrometer software is then reloaded from the
data system. See
“Resetting the Mass Spectrometer”
reset the mass spectrometer.
Three LEDs are located on the power panel: The Pump On LED is green when the rough
pump current sensor detects current to the forepump. The LED is off when the rough-pump
current sensor does not detect current to the forepump. If the current sensor detects a loss of
current when the mass spectrometer is on, the vacuum system vents.
The Vent Valve Closed LED is green whenever the vent valve current sensor detects current
through the vent valve and the vent valve is closed. The LED is off when the vent valve is
open.
!
!
Power In
GC Power Out
Quantum Power Out
A/S Power Out
V~230
50/60 Hz, 2A MAX
V~230
50/60 Hz, 10A MAX
V~230
50/60 Hz, 16A MAX
V~230
Hz, 30A MAX