6
Maintenance
Cleaning Ion Source Components
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
97
5. Disconnect the connectors from the lead pins at the EI/CI Source PCB, lens L1, L2, L3
assembly, and lens L4 assembly. See
Figure 51.
TSQ Quantum GC ion source, installed
6. While holding the ion source assembly, loosen the two thumbscrews that secure the ion
source assembly to the Q0 quadrupole.
7. Pull the ion source assembly away from the Q0 quadrupole and 2 mm to the right to clear
the transfer line bellows.
8. Slide the ion source off the magnet yoke.
9. Place the ion source and magnet yoke on a clean, lint-free cloth.
Q0 quadrupole mount
Lens L1, L2, L3 connector
EI/CI Source PCB
connector
Transfer line
Thumbscrew
Thumbscrew
Lens L4 connector
CAUTION
Burn Hazard
. The ion source might be hot. Allow the ion source to cool
to room temperature before touching it.
Note
Wear clean, lint- and powder- free gloves when you handle the ion source.