6
Maintenance
Disassembling and Reassembling the TSQ Quantum XLS Ion Source Completely
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
107
8. Remove the filament and centering ring from the ion source block.
9. Inspect and install a new filament (
P/N
120320-0030
, in the
TSQ Quantum GC Accessory
Kit
).
a. Verify that the filament wire is centered in the electron lens hole.
b. Insert the filament into the centering ring.
c. Seat the filament on the EI/CI Source PCB.
d. Secure the filament with the retainer clip.
10. Press the heater ring and lens assembly onto the ion source block.
11. Use the retainer clip and threaded standoff to secure the lens assembly to the ion source.
See
12. Reinstall the ion source onto the magnet yoke.
13. Reinstall the ion source into the vacuum manifold. See
.
14. Restart the system. See
“Starting Up the System after a Complete Shutdown”
Disassembling and Reassembling the TSQ Quantum XLS Ion Source
Completely
Disassemble the ion source completely to clean the ion source block or to replace the EI/CI
Source PCB.
See
for the location of the ion source components.
Y
To disassemble the ion source completely
1. Prepare a clean work area by covering the area with lint-free cloth.
Note
Now is a good time to clean the ion source block. Use the procedure described
in
“Cleaning Stainless Steel Parts”
to clean the ion source block.
IMPORTANT
For accurate temperature readings of the heater block, fully screw the
threaded standoff into the ion source to ensure contact between the temperature
sensor with the heater block.
Tools Needed
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Lint-free cloth
Frequency
As needed