4
Tuning and Calibrating
Saving the Tune and Calibration Report
70
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Figure 35.
Tune and calibration report, status page
Results from System Tune and Calibration using FC43 Pos. Ions:
Q1 and Q3 Tuning & Calibrating
10:55:29: Tuning and Calibrating Q1
10:55:39: Tuning Q1MS at mass 69.00 m/z
10:55:41: Opening resolution
10:55:42: Adjusting resolution
10:56:02: The adjusted resolution is -13.33 at the width 0.78
10:56:02: Coarse resolution done - Quick calibration
10:56:07: The adjusted calibration of mass 69.0 is -0.88
10:56:08: Averaging Stopping Curve...
10:56:12: Q1 Quadrupole Offset set to -0.85 for all reswidths
10:56:14: Optimizing Lens 2 for ion 69.00 m/z
10:56:20: Previous Setting = -7.00, New Setting = -0.80
10:56:20: Maximum Intensity = 1.04e+06
10:56:20: 322 % Improvement
10:56:20: Optimizing Lens 4 for ion 69.00 m/z
10:56:26: Previous Setting = -10.00, New Setting = -46.55
10:56:26: Maximum Intensity = 1.61e+06
10:56:26: 60 % Improvement
10:56:26: Optimizing Lens 1-1 for ion 69.00 m/z
10:56:32: Previous Setting = -5.00, New Setting = -0.86
10:56:32: Maximum Intensity = 1.93e+06
10:56:32: 15 % Improvement
10:56:32: Optimizing Lens 1-2 for ion 69.00 m/z
10:56:38: Previous Setting = -21.90, New Setting = -5.00
10:56:38: Maximum Intensity = 2.43e+06
10:56:38: 41 % Improvement
10:56:38: Optimizing Lens 2-1 for ion 69.00 m/z
10:56:45: Previous Setting = -8.45, New Setting = 3.92
10:56:45: Maximum Intensity = 2.56e+06
10:56:45: 17 % Improvement
10:56:45: Tuning Q1MS at mass 501.97 m/s
10:56:47: Opening resolution
10:56:48: Adjusting resolution
10:57:04: The adjusted resolution is -14.38 at the width 0.78
10:57:04: Coarse resolution done - Quick calibration
10:57:10: The adjusted calibration of mass 502.0 is -0.16
10:57:11: Averaging Stopping Curve...
10:57:16: Q1 Quadrupole Offset set to -2.00 for all reswidths
10:57:18: Optimizing Lens 2 for ion 501.97 m/z
10:57:24: Previous Setting = -7.00, New Setting = -2.70
10:57:24: Maximum Intensity = 2.97e+05
10:57:24: 223 % Improvement
10:57:24: Optimizing Lens 1-2 for ion 501.97 m/z
10:57:30: Previous Setting = -50.14, New Setting = -51.11
10:57:30: Maximum Intensity = 4.35e+05
10:57:30: 6 % Improvement
10:57:30: Optimizing Lens 2-1 for ion 501.97 m/z
Page 1 of 10
May 09, 2007 11:17:28
Quantum Tune Master - Tune and Calibration
Signature: __________________________________________