1
Introduction
Scan Modes
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
7
Q1MS and Q3MS Scan Modes
In the Q1MS and Q3MS scan modes, only one stage of mass analysis is performed. The mass
spectrum obtained is equivalent to the mass spectrum obtained from an instrument with a
single mass analyzer. In the one stage of analysis, ions formed in the ion source enter the
analyzer assembly. One of the mass analyzers (Q1 or Q3) is scanned to obtain a complete mass
spectrum. The other rod assemblies (Q2 and Q3, or Q1 and Q2, respectively) act as ion
transmission devices. In the Q1MS scan mode, Q1 is used as the mass analyzer; in the Q3MS
scan mode, Q3 is used as the mass analyzer.
Product Scan Mode
Product scan mode performs two stages of analysis. In the first stage, ions formed in the ion
source enter Q1, which is set to transmit ions of one mass-to-charge ratio. Ions selected by
this first stage of mass analysis are called parent ions. (As a result, Q1 is referred to as the
parent mass analyzer, and the mass-to-charge ratio of ions transmitted by the parent mass
analyzer is referred to as the parent set mass.) Parent ions selected by Q1 then enter Q2, which
is surrounded by the collision cell.
In the second stage of analysis, ions in the collision cell can fragment further to produce
product ions. Two processes produce product ions: by unimolecular decomposition of
metastable ions or by interaction with argon collision gas present in the collision cell. This
latter step is known as collision-induced dissociation (CID). Ions formed in the collision cell
enter Q3 (the product mass analyzer) for the second stage of mass analysis. Q3 is scanned to
obtain a mass spectrum that shows the product ions produced from the fragmentation of the
selected parent ion.
A mass spectrum obtained in the Product scan mode (product mass spectrum) is the mass
spectrum of a selected parent ion.
The Product scan mode is illustrated in
Note
When referring to the first, second, and third rod assemblies as pieces of
hardware, it is convenient to call them Q1, Q2, and Q3, respectively. However, when
discussing their function in MS/MS scan modes, it often adds clarity to refer to them as
the parent mass analyzer, collision cell (ion transmission device surrounded by the
collision cell), and product mass analyzer, respectively.