Replacing the Filament of the TSQ Quantum XLS Ion Source
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
103
See
on page
for the location of the ion source
components.
Y
To replace the ion source filament
1. Prepare a clean work area by covering the area with lint-free cloth.
2. Shut down and vent the TSQ Quantum XLS mass spectrometer. See
3. Remove the ion source. See
“Removing the TSQ Quantum XLS Ion Source”
.
4. Wearing clean, lint- and powder- free gloves, remove the ion source from the magnet
yoke.
Tools Needed
• Filament (P/N 120320-0030)
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Lint-free cloth
Frequency
If filament fails
CAUTION
Shock Hazard
. Unplug the TSQ Quantum XLS mass spectrometer before
proceeding.