6
Maintenance
84
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Gas chromatograph
Replacing the capillary
column
If chromatographic
separation is bad or
if a different
stationary phase is
needed for the
application
Q0 ion optics
Cleaning the Q0
quadrupole and L11
and L12 lenses**
As needed*
Mass analyzer
Cleaning the Q1, Q2,
and Q3 quadrupoles
and lenses**
As needed*
Forepump
Purging the forepump
pump oil
If pump oil is
cloudy
Manufacturer’s
documentation
Forepump
Adding pump oil
If pump oil level is
low
Manufacturer’s
documentation
Forepump
Changing the pump oil Every 4 months or
if pump oil is
cloudy and
discolored
Manufacturer’s
documentation
Ion detection system
Replacing the electron
multiplier assembly**
If noise in
spectrum is
excessive or proper
electron multiplier
gain cannot be
achieved
Electronic modules
Replacing an electronic
module**
If electronic
module fails
PCBs
Replacing a PCB**
If PCB fails
*The frequency of cleaning the components of the mass spectrometer depends on the types and amounts of
samples and solvents that are introduced into the instrument. Cleaning of the Q0, Q1, Q2, and Q3
quadrupoles is rarely (if ever) required.
**A Thermo Fisher Scientific Field Service Engineer must perform this maintenance procedure.
Table 6.
Maintenance procedures (Sheet 2 of 2)
Mass spectrometer
component
Procedure
Frequency
Procedure
location