Using the Direct Sample Probe
Removing the Probe
158
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Figure 77.
Raw file displayed in Qual Browser
Removing the Probe
Y
To remove the probe
1. Withdraw the probe until the guide ball on the probe reaches the first stop on the guide
bar (
2. Close the inlet valve by pushing the lever down.
3. Loosen the inlet valve knob by turning it counter-clockwise.
4. Continue withdrawing the probe completely from the inlet valve by sliding it through the
guide ball track in the guide bar.
CAUTION
Close the inlet valve first. Otherwise, the system vents to the atmosphere.
Do not withdraw the probe beyond the point where the guide ball reaches the first
stop in the guide bar.