5
Changing Ionization Modes
Installing the Ion Volume
82
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
14. Replace the inlet valve plug. Orient the plug to indicate whether an EI or CI ion volume
is installed.
15. Tighten the inlet valve knob clockwise to create a leak-tight seal.
16. (Optional) To prevent the inlet valve from being opened accidentally, pull free and
remove the inlet valve lever.