6
Maintenance
Cleaning Ion Source Components
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
95
6. Reconnect the 3-pin connector to the lead pins on the L1, L2, and L3 lenses. See
7. Reinsert the capillary column into the ion source.
a. Using the I/R tool, remove the ion volume. See
b. Push the capillary column in until you can see it in the ion source.
c.
Pull the column back just far enough that you cannot see it.
d. Tighten the transfer line nut and transfer line union.
e.
Using the I/R tool, install the ion volume. See
8. Close and secure the cover to the ion source vacuum manifold.
Cleaning the Lenses of the TSQ Quantum GC Ion Source
If cleaning the ion volume does not restore system performance, clean the ion source lenses.
Y
To clean the ion source lenses
1. Shut down and vent the system. See
“Shutting Down the System Completely”
2. Remove the ion source. See
“Removing the TSQ Quantum XLS Ion Source”
3. Remove the ion source lens assembly from the ion source and disassemble it. See
“Removing the L1, L2, L3 Lens Assembly and the Q00 Quadrupole Lens Assembly”
4. Clean the lenses and the spacers. See
“Cleaning Stainless Steel Parts”
and
“Cleaning Non-Stainless Steel or Hybrid Part”
, respectively.
5. Reassemble the ion source lens assembly and install it onto the ion source. See
“Reassembling the L1, L2, L3 Lens Assembly”
.
Note
Lens L4 requires cleaning less often than lens L1, L2, and L3.
CAUTION
Shock Hazard
. Unplug the TSQ Quantum GC before proceeding.
CAUTION
Burn Hazard. The ion source might be hot. Allow the ion source to cool to
room temperature before touching it.