6
Maintenance
Cleaning Ion Source Components
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
91
9. Place the ion source and magnet yoke on a clean, lint-free cloth.
10. Close the lid of the ion source vacuum manifold.
Removing the L1, L2, L3 Lens Assembly and the Q00 Quadrupole Lens Assembly
The ion source includes the L1, L2, L3 lens assembly and the Q00 quadrupole lens assembly.
You must remove and disassemble the L1, L2, L3 lens assembly to clean the lenses. A retainer
clip fastens the L1, L2, L3 lens assembly and the Q00 quadrupole lens assembly to the ion
source block.
To remove the ion source lens assemblies
1. Wearing clean, lint- and powder- free gloves, unscrew the threaded hexagonal standoff
and remove the retainer clip that secures the lens assembly to the ion source block. See
2. Remove the Q00 quadrupole lens assembly and the L1, L2, L3 lens assembly from the
heater cartridges of the ion source. Place them on a clean surface.