4
Tuning and Calibrating
Saving the Tune and Calibration Report
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
71
Figure 36.
Tune and calibration report, parameter optimization plots
Y
To save the tune and calibration report
1. In the System Tune and Calibration dialog box, click
Save Report
. The Save As dialog
box appears.
2. Enter the name and the path of the report, and click
Save
.
Optimizing Lens 2 for Q1MS
-50
-40
-30
-20
-10
0
Lens 2 Voltage (V)
0
20
40
60
80
100
y
ti
s
n
e
t
nI
e
vi
t
al
e
R
May 9, 2007
Previous Setting
Optimum Setting
Mass 69.00m/z : 322 % Improvement
Mass 501.97m/z : 223 % Improvement
Mass 218.99m/z : 94 % Improvement
Optimizing Lens 2 for Q3MS
-50
-40
-30
-20
-10
0
Lens 2 Voltage (V)
0
20
40
60
80
100
y
ti
s
n
e
t
nI
e
vi
t
al
e
R
May 9, 2007
Previous Setting
Optimum Setting
Mass 69.00m/z : 235 % Improvement
Mass 501.97m/z : 281 % Improvement
Mass 218.99m/z : 57 % Improvement
Optimizing Lens 4 for Q1MS
-50
-40
-30
-20
-10
0
Lens 4 Voltage (V)
0
20
40
60
80
100
y
ti
s
n
e
t
nI
e
vi
t
al
e
R
May 9, 2007
Previous Setting
Optimum Setting
Mass 69.00m/z : 60 % Improvement
Optimizing Lens 1-1 for Q1MS
-10
-8
-6
-4
-2
0
Lens 1-1 Voltage (V)
0
20
40
60
80
100
y
ti
s
n
e
t
nI
e
vi
t
al
e
R
May 9, 2007
Previous Setting
Optimum Setting
Mass 69.00m/z : 15 % Improvement
Optimizing Lens 1-1 for Q3MS
-10
-8
-6
-4
-2
0
Lens 1-1 Voltage (V)
0
20
40
60
80
100
y
ti
s
n
e
t
nI
e
vi
t
al
e
R
May 9, 2007
Previous Setting
Optimum Setting
Mass 69.00m/z : 6 % Improvement
Optimizing Lens 1-2 for Q1MS
-250
-200
-150
-100
-50
0
Lens 1-2 Voltage (V)
0
20
40
60
80
100
y
ti
s
n
e
t
nI
e
vi
t
al
e
R
May 9, 2007
Previous Setting
Optimum Setting
Mass 69.00m/z : 41 % Improvement
Mass 501.97m/z : 6 % Improvement
Mass 218.99m/z : No Improvement
Quantum Tune Master - Tune and Calibration