Tuning and Calibrating
Running Auto Tune and Calibration
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
67
Figure 33.
System Tune and Calibration dialog box
3. In the Compound list, select
FC43 pos. ions
. This automatically selects the positively
charged FC-43 ions to be used for automatic tuning and calibrating.
4. Select
Tune A
as the tune algorithm.
5. Click
Start
to start the automatic tuning and calibration procedure.
The Status box displays real-time messages about the system tune and calibration so that
you can monitor the progress of each sub-procedure. After a sub-procedure is complete,
the result is reported (for example, whether it passed or failed). At the end of the entire
procedure, it displays a summary.
• If errors occur during the automatic tuning and calibration procedure, go to
• If the automatic tuning and calibration procedure finishes without errors, go
6. If errors occur during the automatic tuning and calibration procedure, restore the
previous mass spectrometer device settings and perform the tuning and calibration
procedure again by completing the following steps:
a. To restore the prior tuning and calibration settings, click
Undo
.
b. To reload the prior tuning and calibration settings to the mass spectrometer, click
Accept
.
c.
Troubleshoot and correct the problem that caused the tuning and calibration
procedure to fail. See
and restart the tuning and calibration procedure.
Note
Tune A is the first choice of tune algorithm if there is a regular EI or CI ion
volume installed in the ion source. Tune B is the first choice of tune algorithm if there
is a closed EI ion volume installed.