Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
147
8
Using the Direct Sample Probe
With the direct sample probes (direct insertion probe and direct exposure probe) you can
introduce compounds directly into the ion source without GC column separation. This
chapter covers, step-by-step, a direct insertion probe (DIP) experiment using
perfluorotetracosane (CF
3
(CF
2
)
22
CF
3
, molecular weight 1238.18, melting point 190
ºC
).
You can perform a similar experiment using your compound.
P
Creating an Instrument Method
An instrument method contains the settings that the TSQ Quantum XLS or TSQ Quantum
GC mass spectrometer uses during an experiment. Use the Instrument Setup view to create
the instrument method.
Y
To create an instrument method
1. On the Xcalibur Home Page, click the
Instrument Setup
icon to open the Instrument
Setup view.
2. Click the
TSQ
icon in the viewbar.
3. Choose
TSQ > Regular Method
to open the regular method Scan Editor page.
4. On the Scan Editor page, create a one segment, one scan event instrument method. In
this example, the segment is 8 minutes long (see
Contents
•
•
•
Preparing the Probe and Inlet Valve
•
Preparing the Mass Spectrometer
•
•
Examining the Raw Data in Qual Browser
•