Using the Direct Sample Probe
Creating a Sequence
150
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
b. Save the instrument method as a .meth file. In this example we name the instrument
method
Dip test.meth
.
Creating a Sequence
A sequence contains sample information—one sample per row. In this example the sequence
has one sample.
Y
To create a sequence
1. On the Xcalibur Home Page, click the
Sequence Setup
icon to open the Sequence Setup
view (see
2. Enter the name and path of the instrument method. In this example, it is
C:\Xcalibur\methods\DIP test.
3. Enter a file name and path for the raw file that contains the acquired data. In this
example, it is C:\Xcalibur\Data\data01.
Figure 67.
Sequence Setup view, showing one sample in the sequence
Note
If the sequence has more than one sample, Xcalibur prompts you to reload the
probe after each sample.