6
Maintenance
Disassembling and Reassembling the TSQ Quantum GC Ion Source Completely
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
115
Cleaning Non-Stainless Steel or Hybrid Part
Y
To clean non-stainless steel or hybrid parts
1. Remove contamination from stainless steel surfaces. It is only necessary to clean a surface
that comes in contact with the ion beam.
a. Use a slurry of number 600 aluminum oxide in glycerol and a cleaning brush or
cotton-tipped applicator. Contamination appears as dark or discolored areas, but
often is not visible.
b. Clean each part thoroughly, even if no contamination is visible.
c.
To clean the inside corners, use the wooden end of an applicator cut at an angle.
d. Use a Dremel tool with the polishing swab at its lowest speed to increase cleaning
efficiency, as well as decrease the time required to clean the items. To prevent personal
injury, be sure to keep the tool away from possible hazards, such as standing water or
flammable solvents.
Tools Needed
• Acetone, reagent grade (or other suitable polar solvent)
• Aluminum oxide abrasive, number 600 (P/N 32000-60340)
• Applicators, cotton-tipped (P/N 00301-01-00015)
• Beaker, 450 mL
• Clean, dry gas
• De-ionized water
• Detergent (Alconox, Micro, or equivalent)
• Dremel rotary tool or equivalent (recommended)
• Forceps (P/N 76360-0400)
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Glycerol, reagent grade
• Lint-free cloth
• Protective eyewear
• Tap water
• Toothbrush, soft
Frequency
As needed to clean non-stainless steel parts (such as aluminum, ceramic, or
gold-plated), or to clean hybrid parts that are partially made of stainless steel:
• Lens holder and spacers
• Filament spacer
• Heater ring
CAUTION
Material and Eye Hazard.
Wear impermeable laboratory gloves and eye
protection when performing cleaning procedures.