SOLA II User Guide 9-1
Chapter 9
At Line-Grab Sample
This function was developed primarily for pipeline applications. The sampling
system was designed for the pipeline enclosure in order to provide an
additional pump to facilitate introduction of sample. Access this top-level
menu to enable the analyzer to analyze and report the results of a sample
previously collected from a source other than the regular process input. The
sample to be analyzed is introduced into the instrument in the same manner
that a calibration standard is.
There are three parameters to configure: starting sample ID, average sampling
time, and the relative standard deviation of the samples taken.
1.
Each sample is identified with a number (starting sample ID). The
numbers increment sequentially until a new starting sample ID is
programmed.
Figure 9–1.
2.
Average sampling time is the number of seconds that the sample is
averaged to compute the result. It can be between 60 and 240 seconds.
Figure 9–2.