![Thermo Scientific SOLA II User Manual Download Page 118](http://html1.mh-extra.com/html/thermo-scientific/sola-ii/sola-ii_user-manual_1105272118.webp)
Watlow EZ-ZONE® PM6 Series
Software Configuration
A-4 SOLA II User Guide
Thermo Fisher Scientific
Parameter
Default
Comment
Display
Software
Display
Software
Sensor Type
Thermocouple
TC Linearization
S
For Type S thermocouple
Filter
0.5
For smoothing the process signal.
Input Error Latching
Off
Display Precision
Whole
Table A–3.
Pyrolyzer settings: Setup Page > Control Loop 1
Parameter
Default
Comment
Display
Software
Display
Software
Heat Algorithm
PID
The method of heat control.
Cool Algorithm
Off
TRU-TUNE+ Enable
No
Autotune Aggressiveness
Critical
Critical damped aggressiveness of
autotuning calculations.
User Failure Action
User
When user switches to manual
mode, controller sets output power
to last open-loop set point entered
by user.
Input Error Failure
Off
When an input error switches
control to manual mode, controller
sets output to 0%.
Open Loop Detect Enable
No
Ramp Action
Off
Low Set Point
0°C
The minimum value of the closed
loop set point range.
High Set Point
1150°C
The maximum value of the closed
loop set point range.
Set Point Open Limit Low
-100.0%
The minimum value of the open
loop set point range.
Set Point Open Limit High
100.0%
The maximum value of the open
loop set point range.
Table A–4.
Pyrolyzer settings: Setup Page > Output 1
Summary of Contents for SOLA II
Page 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Page 4: ......
Page 6: ......
Page 12: ......
Page 16: ......
Page 38: ......
Page 50: ......
Page 84: ......
Page 92: ......
Page 150: ......
Page 160: ......
Page 168: ......
Page 188: ......
Page 194: ......
Page 198: ......
Page 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...