Startup & Shutdown
Emergency Shutdown
Thermo Fisher Scientific
SOLA II User Guide 4-5
3.
Allow the instrument to cool.
a.
If the analyzer is used in an ATEX hazardous area with a T3 rating, do
not open the oven door for 45 minutes, 50 minutes for Zone 2 unit.
b.
If the analyzer is used in an ATEX hazardous area with a T4 rating,
do not open the oven door for 140 minutes.
Caution
Failure to allow adequate cooling time before opening the oven can
lead to equipment damage or injury to personnel.
σ
4.
Turn off all air supplies.
Caution
Parts of the instrument may be hot even after power is removed.
Allow the system to cool completely before performing maintenance.
σ
1.
Close the sample supply to the system.
2.
Turn off the main power to the system.
Emergency
Shutdown