Configuration
Calibration
Thermo Fisher Scientific
SOLA II User Guide 6-17
The values and a graph for a typical calibration are shown below.
Low cal value = 0.00 ppm
Low cal = 555.600 kHz
High cal value = 100.00 ppm
High cal = 15000.000 kHz
Figure 6–44. Typical SOLA II calibration
The high and low calibration can be performed independently. Changing
either one of the calibration points will change the slope of the line. This is
illustrated in the next two graphs.
Figure 6–45. High calibration