Watlow EZ-ZONE® PM6 Series
Software Configuration
Thermo Fisher Scientific
SOLA II User Guide A-13
Parameter
Default
Comment
Set Point Open Limit High
100.0%
The maximum value of the open
loop set point range.
Table A–19.
Oven settings: Setup Page > Output 1
Parameter
Default
Comment
Display
Software
Display
Software
Output Function
Heat
Output Function Instance
1
Output Control
Fixed Time Base
Output Time Base
0.5
Output Low Power Scale
0%
The power output will never be
less than the value specified and
will represent the value at which
output scaling begins.
Output High Power Scale
70%
The power output will never be
greater than the value specified
and will represent the value at
which output scaling stops.
Table A–20.
Oven settings: Setup Page > Output 2
Parameter
Default
Comment
Display
Software
Display
Software
Output Function
Alarm
Output Function Instance
1
Table A–21.
Oven settings: Setup Page > Alarm 1
Parameter
Default
Comment
Display
Software
Display
Software
Alarm Type
Deviation Alarm
Alarm Hysteresis
2°C
Alarm Logic
Open On Alarm
The output condition during an
alarm state.
Alarm Sides
Both
Both high and low sides can
trigger the alarm.
Summary of Contents for SOLA II
Page 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Page 4: ......
Page 6: ......
Page 12: ......
Page 16: ......
Page 38: ......
Page 50: ......
Page 84: ......
Page 92: ......
Page 150: ......
Page 160: ......
Page 168: ......
Page 188: ......
Page 194: ......
Page 198: ......
Page 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...