Maintenance & Troubleshooting
Replacing the Pyrolyzer Heater
Thermo Fisher Scientific
SOLA II User Guide 10-7
Figure 10–3. Remove the pyrolyzer housing lid
8. Gently lift the lid off and set it aside.
Note
The pyrolyzer tube, insulation, and Kaowool insulation will be
attached to the lid (Figure 10-4). Carefully lift all of it out to avoid
damaging the pyrolyzer tube.
σ
Figure 10–4. Housing lid with insulation and pyrolyzer tube attached
9. As shown in Figure 10-5, the heater wires are attached to the terminal
block. Remove the heater wires from the terminal block and straighten
the wires. Remove the fiberglass insulation that covers the heater wires
(Figure 10-6). Set the insulation aside. You will need it later.