Configuration
Alarms Setup
Thermo Fisher Scientific
SOLA II User Guide 6-15
7.
Select one of the TS Level Alarms modes. Select Disable, Non Latching, or
Latching mode for the concentration alarms. If you select nonlatching, the
alarm clears when the measured concentration drops below the associated
value. If you select latching, the alarm must be acknowledged before it will
clear.
Figure 6–40.
8.
Set the High Alarm for stream 1 by entering the first level of high
concentration for stream 1 at which the alarm should activate.
Figure 6–41.
9.
Set the High High alarm for stream 1 by entering the second level of high
concentration for stream 1 at which the alarm should activate. There are
also submenus for setting the high alarm and the high high alarm for
stream 2. If applicable, set these alarms as done for stream 1.
Figure 6–42.