Configuration
Dual Range Setup
6-8 SOLA II User Guide
Thermo Fisher Scientific
This menu allows you to set the dual range mode.
1.
Set the dual range mode to Remote Range Select or to Auto Range for the
4–20 mA output signal. If you select Auto Range, the 4–20 mA output
signal switches to high range when the reading goes above the low range
full scale value. When the signal drops below 90% of the low range full
scale value, the system switches back to low range.
When Remote Range Select is enabled, output scaling is controlled by a
remote contact input. When the input contacts are open, the system
operates on low range. When the input contacts are closed, the system
operates on high range.
Figure 6–20.
Figure 6–21.
2.
The submenu shown below enables you to specify the full scale value (20
mA) for the high range of stream 1 as output on the 4–20 mA signal.
Figure 6–22.
Dual Range
Setup