The X-Purge System
X-Purge Specifications
Thermo Fisher Scientific
SOLA II User Guide G-3
Table G–1.
General specifications
Certifications
CSA with C and US Mark: Class 1, Div. 1, Groups B, C, D
hazardous areas
ATEX: Zone 1, Ex px II T
IECEx: Ex d[p] IIC T4
Programming
Programmable time delay
Function
Monitors 2 pressurized enclosures and 2 exhaust port
flows
Table G–2. Normal conditions
Normal conditions
Power
AC applied to X-Purge unit
Switches
Normal/Bypass: Set to NORMAL
Time delay relay: Set to number of seconds for delay;
time delay typically set from 480 to 600 seconds (see
application notes for individual system)
X-Purge pressure
Minimum 10 psig
Y-Purge pressure (analyzer
purge)
Set to approximately 10 psig or as specified in the
application notes shipped with the analyzer
Oven air pressure
Minimum 10 psig or as specified in instrument application
notes or system log books
Oven door
Closed tightly
Electronic housing door
Closed tightly
Instrument air ball valve for
X-Purge unit
Open
Table G–3. Utility requirements
Utility requirements
Instrument air
60–80 psig, 8 SCFM (maximum)
Instrument air quality
Water and oil free, -40°C (-40°F) dew point, particles <
5µ, ISA grade, hydrocarbon free
AC power
110 Vac, 50/60 Hz
X-Purge
Specifications
Summary of Contents for SOLA II
Page 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Page 4: ......
Page 6: ......
Page 12: ......
Page 16: ......
Page 38: ......
Page 50: ......
Page 84: ......
Page 92: ......
Page 150: ......
Page 160: ......
Page 168: ......
Page 188: ......
Page 194: ......
Page 198: ......
Page 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...