Rotary Valve Service
Two-Position Air Actuator O-ring Replacement
B-6 SOLA II User Guide
Thermo Fisher Scientific
You will need the following to perform the tasks described in this section.
λ
9/64” hex driver
λ
3/8” open-end wrench
λ
3/16” screwdriver
λ
An awl or small jeweler’s screwdriver
λ
Silicone lubricant (such as Dow Corning
DC-111)
λ
Lint-free tissues and a clean shop rag
λ
Standard O-ring kit (VICI P/N OR)
λ
High temperature O-ring kit (VICI P/N ORT)
1.
Apply air pressure to the actuator inlet nearest the valve. Then use the
open-end wrench to remove the air supply lines from the actuator.
Figure B–5.
2.
Remove the valve and valve-mounting hardware from the actuator (as
shown in Figure B–5):
a.
Use the 9/64" hex driver to loosen the HWSC-SC8-10B/socket-head
screw in the black anodized CR3/clamp ring on the actuator.
b.
Pull off the standoff with the valve attached.
c.
Use the 9/64” hex driver to remove the two HWSC-SC8-6/socket-head
screws that hold the clamp ring to the actuator.
Two-Position Air
Actuator O-ring
Replacement
Disassembly
Summary of Contents for SOLA II
Page 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Page 4: ......
Page 6: ......
Page 12: ......
Page 16: ......
Page 38: ......
Page 50: ......
Page 84: ......
Page 92: ......
Page 150: ......
Page 160: ......
Page 168: ......
Page 188: ......
Page 194: ......
Page 198: ......
Page 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...