Configuration
Inject Setup
Thermo Fisher Scientific
SOLA II User Guide 6-9
3.
The submenu shown below enables you to specify the full scale value (20
mA) for the low range of stream 1 as output on the 4–20 mA signal.
Figure 6–23.
Set the 20 mA scaling for the high and low ranges for stream 2 using the
submenus for stream 2.
1.
Enter the Inject Setup menu to set injection parameters.
Figure 6–24.
2.
The injection time is the amount of time that the injection valve stays in
the injection position. Set the inject time at half of the inject rate (1 to
9999 seconds).
Figure 6–25.
Inject Setup