Configuration
Analyzer Setup
6-2 SOLA II User Guide
Thermo Fisher Scientific
access the menus. Change the passcode to prevent unauthorized access to
the menus.
Figure 6–3.
3.
Return to the Analyzer Setup menu. Press the down arrow until the
Remote Control submenu appears (Figure 6–4), allowing you to
enable/disable the remote control function.
in Chapter 3 addresses which functions may be operated remotely.
Figure 6–4.
4.
The Average Time submenu allows you to set the averaging time. The
average time is a period (1–240 seconds) during which the analyzer takes
SO
2
measurements used to determine a moving average of the results. For
example, if you set the averaging time to 10 seconds, the average
concentration of the last 10 seconds is output at each update. If you set
the averaging time to 60 seconds, the average concentration of the last 60
seconds is output at each update. Thus, lower averaging time means faster
response to concentration changes. Longer averaging times are typically
used to smooth output data.
See the figure below.
Figure 6–5.