The Transmission indicator lights when transmission mode is
selected. The Reflection indicator lights when reflection mode is
selected.
4. Rotate the round visible polarizer with your finger until the
field of view becomes dark.
The polarizers are now crossed.
5. Install the sample and focus.
If you have difficulty seeing the sample to focus, try rotating the
sample.
If you still cannot see the sample after completely rotating it, it is
likely that there is no polarization data. Try another technique.
6. While viewing the sample through the eyepieces or through
the video display, rotate the sample or the round visible
polarizer to see the differences in sample composition.
Note
The optional rotatable stage is ideal for this type of experiment. The
universal slide holder can also be rotated. We recommend that you
rotate the sample rather than the polarizer while looking for
differences in sample composition.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...