Continuµm User Guide 125
Also, be sure that the beamsplitter and source installed in your
spectrometer are compatible with the microscope detector you have
selected. The table that follows lists the beamsplitter-detector
combinations that can be used.
Beamsplitter
Near-IR
Mid-IR*
Far-IR
Detector
Quartz
CaF
2
XT-KBr
KBr
CsI
Solid
substrate
MCT-A*
OK
OK
Best
Best
OK
X
MCT-A
OK
OK
Best
Best
OK
X
MCT-B
OK
OK
Best
Best
OK
X
InGaAs
OK
Best
OK
X
X
X
Notes:
Best = Optimum beamsplitter-detector combination.
OK = Compatible beamsplitter-detector combination.
X
= Incompatible beamsplitter-detector combination.
* A ZnSe beamsplitter may also be used as a mid-IR beamsplitter in extremely humid environments.
The next table lists the spectral ranges of compatible beamsplitter-
detector-source combinations.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...