Continuµm User Guide
65
4. Start OMNIC and select “Transmission discrete point –
standard sample“ from the Experiment drop-down list box.
5. Choose Show Atlµs Window from the Atlµs menu.
When the Atlµs window opens you will see the video pane on
the right side of the screen, the navigation (stage position) pane
on the left, and the Atlµs tool palettes near the bottom of the
screen. You may or may not be able to see a clear image in the
video pane at this point.
You may also be able to see the default background (crosshairs
with the letter B) and sampling points (crosshairs with a number
and an asterisk). Whenever video capture is enabled at the
sampling point, an asterisk appears in the designator. When more
than one discrete sampling point is mapped for an experiment,
data will be collected from sampling point 1, then 2, and so on.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...