Continuµm User Guide
67
7. Use the joystick to move the sample into the field of view and
then sharply focus on the XL pattern.
You should now be able to see a clear image of the ink in the
video pane.
Note
A yellow circle in the Bench Status field usually indicates that the
detector has not been cooled. If you already added liquid nitrogen to
the detector dewar, simply wait for temperature to stabilize. If you
have not yet cooled the detector, do so before attempting to collect
data.
You may also see a yellow circle in the Bench Status field if the
quality checks, available through the Quality tab in Experiment
Setup have not been disabled. The default quality check settings
were designed for experiments conducted in the spectrometer sample
compartment. They are not useful for sampling with the
microscope.
8. Set the irises.
Fully open both field irises.
Summary of Contents for Nicolet Continuum
Page 1: ......
Page 9: ...Continuµm User Guide 5 Front panel ...
Page 10: ...Thermo Scientific 6 Control panel on left side ...
Page 45: ...Continuµm User Guide 41 2 Turn on Reflex aperture illuminator ...
Page 48: ...Thermo Scientific 44 ...
Page 58: ...Thermo Scientific 54 Adjust the reflection aperture iris for good contrast in the video image ...
Page 97: ...Continuµm User Guide 93 2 Lower the condenser completely Use the condenser focus knob ...
Page 98: ...Thermo Scientific 94 3 If the universal slide holder is in place remove it from the stage ...
Page 114: ...Thermo Scientific 110 2 Lower the condenser fully Use the condenser focus knob ...
Page 115: ...Continuµm User Guide 111 3 If the universal slide holder is in place remove it from the stage ...
Page 126: ...Thermo Scientific 122 ...
Page 154: ...Thermo Scientific 150 ...